Página 1 - Ultraview 1050 Monitors

Ultraview® 1030 and Ultraview 1050 Monitors90367, 90369Service Manual070-0706-01 Rev. K

Página 2

Ultraview 1030/1050 Monitors — Service Manual2-2Figure 2-1: Unpacking the Ultraview 1030/1050 monitorBatteriesEach monitor is supplied with one VCR-ty

Página 3 - Table of Contents

Ultraview 1030/1050 Monitors — Service Manual5-1201030100 Diagnostics Port Test Failed Replace CPU PCBA.01030200 860 Test Failed Replace CPU PCBA.0103

Página 4

Ultraview 1030/1050 Monitors — Service Manual5-13Troubleshooting01030700 PCI Bridge Test Failed Replace CPU PCBA.01030701PCI Bridge configuration not

Página 5 - Introduction

Ultraview 1030/1050 Monitors — Service Manual5-1401030B06 IRTS uP RAM failed Replace CPU PCBA.01030B07 IRTS uP ROM failed Replace CPU PCBA.01030B08 IR

Página 6 - Monitor Hardware Options

Ultraview 1030/1050 Monitors — Service Manual5-15TroubleshootingSystem TroubleshootingThis section describes troubleshooting procedures which can be u

Página 7

Ultraview 1030/1050 Monitors — Service Manual5-16Caution:• When troubleshooting, partial disassembly may be required. Ensure that the external power s

Página 8

Ultraview 1030/1050 Monitors — Service Manual5-17TroubleshootingModule Slot Troubleshooting1 Independently verify the module’s operation on another un

Página 9 - Unpacking

Ultraview 1030/1050 Monitors — Service Manual5-184 Verify that the wide, flat flex cable connecting the wireless card to the CPU PCBA is fully inserte

Página 10 - Batteries

Ultraview 1030/1050 Monitors — Service Manual5-19TroubleshootingAudio Line Output Troubleshooting1 Independently verify the operation of the equipment

Página 11 - Sealed Lead Acid (SLA)

Ultraview 1030/1050 Monitors — Service Manual5-20Battery Fuel GaugeWhen the monitor is powered by batteries only, with 1 or 2 batteries installed, a b

Página 12 - Caution:

Ultraview 1030/1050 Monitors — Service Manual5-21TroubleshootingFigure 5-9: Location of D2 on recorder CPUIf a diagnostic error occurs, the recorder w

Página 13 - Assembly

Ultraview 1030/1050 Monitors — Service Manual2-3SetupNickel Metal Hydride (NiMH)NiMH batteries are shipped in a discharged state and will require some

Página 14 - Back Panel Connections

Ultraview 1030/1050 Monitors — Service Manual5-22During normal operation, D2 will be dimly lit (it is actually flashing very rapidly). LED Error Codes

Página 15

Ultraview 1030/1050 Monitors — Service Manual5-23Troubleshooting14 Stuck I/O bus bit 4 low RCDIAG15 Stuck I/O bus bit 5 low RCDIAG16 Stuck I/O bus bit

Página 16 - Power-ON Test

Ultraview 1030/1050 Monitors — Service Manual5-2426 EEPROM Tests - New checksum no good RCDIAG38Printer Tests - WRRDY inactive or -SYNC active after r

Página 17

Ultraview 1030/1050 Monitors — Service Manual5-25TroubleshootingCapnography TroubleshootingThe capnograph option requires the following supportive equ

Página 18 - Maximum Cable Lengths

Ultraview 1030/1050 Monitors — Service Manual5-26The following errors are indicated by a flashing code:If the Channel does not sign ON:1 Check that si

Página 19 - Software Configuration

Ultraview 1030/1050 Monitors — Service Manual5-27TroubleshootingOxygen Preamp/Analog to Digital Converter ProblemAttach a known good O2 sensor and cab

Página 20

Ultraview 1030/1050 Monitors — Service Manual5-28Problem MessagesIf the following error messages are displayed for the Capnograph Option (G or H), per

Página 21

Ultraview 1030/1050 Monitors — Service Manual5-29Troubleshooting6 If the loopback test does not pass, replace the CO2/SDLC PCBA.7 If all the voltages

Página 23 - Keyboard Menu

Contents6-1PartsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Field

Página 24 - Biomed Setup Keys

Ultraview 1030/1050 Monitors — Service Manual2-4After long-term storage, the battery delivers less than full capacity on the first cycle. When a batte

Página 25 - NODE DIRECTORY Key

Ultraview 1030/1050 Monitors — Service Manual6-2Pin, hinge, battery door 214-0318-xxBattery eject 214-0518-xxSpring, extension 1.5 inches, 0.25 od 214

Página 26 - RECORDER SETUP Key

Ultraview 1030/1050 Monitors — Service Manual6-3Parts90367/90369-G or -H (Capnography)Option Description Part NumberGCable assy, sensor interface to C

Página 27 - SELECT RECORDERS Key

Ultraview 1030/1050 Monitors — Service Manual6-490310 Description Part NumberClip, flat cable 344-0028-xxPlate, shoe, 3 inch, module hsg, 90485 386-01

Página 28 - Select beds/subnets

Ultraview 1030/1050 Monitors — Service Manual6-5PartsDrawingsTitleDrawing Part NumberDrawing NumberTop Assembly 653-0056-00 1 (4 sheets)I/O Bay, Capno

Página 30 - HIGH MEDIUM

ContentsA-1Appendix A — GlossaryTerms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 31 - TONE CONFIGURATION Menu Tree

Ultraview 1030/1050 Monitors — Service ManualA-2DB-26“D” shaped 26-pin connector of either male or female gender.DB-9“D” shaped 9-pin connector of eit

Página 32 - SERVICE FUNCTIONS Menu

Ultraview 1030/1050 Monitors — Service ManualA-3Appendix A — GlossaryHigh level outputAnalog signals supplied through a separate connector for use wit

Página 33 - STATISTICS

Ultraview 1030/1050 Monitors — Service ManualA-4PCIS™Patient Care Information SystemPCMS™Patient Care Management SystemPFAILPower Failure notification

Página 34 - RESET MONITOR Key

Ultraview 1030/1050 Monitors — Service ManualA-5Appendix A — GlossaryRTGLReal time graphics libraryRTS Ready-to-send signal used in communications pro

Página 35 - Network Configuration

Ultraview 1030/1050 Monitors — Service Manual2-5SetupAssemblyOpen the battery compartment door and insert the battery(ies) into the battery compartmen

Página 36 - Module Tests

Ultraview 1030/1050 Monitors — Service ManualA-6Transceiver (Ethernet)Device located on coax cable or line powered attaching monitors to the network.

Página 37 - 90310 Front Panel

ContentsB-1Appendix B — Electromagnetic CompatibilityElectromagnetic Emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 38

Ultraview 1030/1050 Monitors — Service ManualB-2Electromagnetic ImmunityNote:This equipment is intended for use in the electromagnetic environment spe

Página 39

Ultraview 1030/1050 Monitors — Service ManualB-3Appendix B — Electromagnetic CompatibilityFrequency Separation DistancesNote:This equipment is intende

Página 40

Ultraview 1030/1050 Monitors — Service ManualB-4InterferenceRF fields pulsating at a sustained rate within the physiological pass band of this equipme

Página 41

C-1Appendix C — SymbolsThe following list of international and safety symbols describes all symbols used on Spacelabs Medical products. No one product

Página 42 - Optional Capnography

Ultraview 1030/1050 Monitors — Service ManualC-2Standby STANDBY KeyPAUSE or INTERRUPT Slow RunReset Power Indicator LEDAlarmTemporary Shut Off of Alar

Página 43 - 90367/90369 Monitor

Ultraview 1030/1050 Monitors — Service ManualC-3Appendix C — SymbolsData Input/Output Input/OutputInput ResetMenu Keys Waveform/Parameter KeysMonitor

Página 44 - CPU PCBA

Ultraview 1030/1050 Monitors — Service ManualC-4Open Padlock Closed PadlockDown Arrow Up ArrowHard Drive Power Indicator LEDAntenna Mermaid ConnectorM

Página 45 - MPC860-PCI Bridge

Ultraview 1030/1050 Monitors — Service ManualC-5Appendix C — SymbolsCaution - hazardous voltages. To reduce risk of electric shock, do not remove the

Página 46 - ISA Bridge

Ultraview 1030/1050 Monitors — Service Manual2-6Back Panel ConnectionsConnections for the external power supply and the optional SDLC network and Remo

Página 47 - ISA I/O Buffers

Ultraview 1030/1050 Monitors — Service ManualC-6IEC 60601-1 Type CF equipment. The unit displaying this symbol contains an F-type isolated (floating)

Página 48 - Clock Distribution

Ultraview 1030/1050 Monitors — Service ManualC-7Appendix C — SymbolsAbbreviations used as symbols are shown below.Latex-free ReusableRadio transmitti

Página 49 - Hardware Reset

Ultraview 1030/1050 Monitors — Service ManualC-8EMGemgElectromyogram ESISElectrosurgical Interference SuppressionEXT External FECG Fetal Electrocardio

Página 50 - Interrupts

Ultraview 1030/1050 Monitors — Service Manual2-7SetupInstalling the Optional Recorder AssemblyCaution:Unless purchased separately for a monitor alread

Página 51

Ultraview 1030/1050 Monitors — Service Manual2-8Inserting Recorder PaperTwo rolls of recorder thermal paper have been included with the recorder assem

Página 52 - Power Supply System

Ultraview 1030/1050 Monitors — Service Manual2-9SetupCablingVarious cables and adapters are provided for interconnecting 90367/90369 monitor component

Página 53

Ultraview 1030/1050 Monitors — Service Manual2-10Cables and AdaptersRefer to Parts on page 6-1 for additional cable part numbers.Maximum Cable Lengths

Página 54 - Battery Chargers

Ultraview 1030/1050 Monitors — Service Manual2-11SetupEthernet CableBulk CablesIf desired, cables can be constructed from bulk 500 foot (152.4 meters)

Página 55

©2005 Spacelabs Medical, Inc.All rights reserved. Contents of this publication may not be reproduced in any form without the written permission of Spa

Página 56 - Boot Sequence Overview

Ultraview 1030/1050 Monitors — Service Manual2-122 Enter the correct case-sensitive password to display the Biomed Monitor Setup menu:SETUPPRESELECTED

Página 57

Ultraview 1030/1050 Monitors — Service Manual2-13SetupSETUPPRESELECTEDRECORDINGSRECORDERSETUPALARM USERACCESSUNITS OFNETWORK SERIALPORTS SETUP MEASURE

Página 58 - Touchscreen PCBA

Ultraview 1030/1050 Monitors — Service Manual2-14SETUPPRESELECTEDRECORDINGSRECORDERSETUPALARM USERACCESSUNITS OFNETWORK SERIALPORTSSETUP MEASUREMOREER

Página 59 - 26 gnd Input

Ultraview 1030/1050 Monitors — Service Manual2-15SetupKeyboard MenuThe Keyboard menu appears whenever a menu is selected that requires entering alphan

Página 60 - I/O Bay PCBA

Ultraview 1030/1050 Monitors — Service Manual2-16Biomed Setup KeysNETWORK SETUP KeyThe Network Setup key enables you to define a subnet name, node ID,

Página 61

Ultraview 1030/1050 Monitors — Service Manual2-17SetupTCP/IP Setup KeyTouch TCP/IP SETUP to display the Keyboard menu (refer to Figure 2-6 on page 2-1

Página 62 - Recorder Interface PCBA

Ultraview 1030/1050 Monitors — Service Manual2-18SUBNET ACCESS KeyThe SUBNET ACCESS key limits this monitor’s accessibility to beds physically located

Página 63 - ICB (Interconnect PCBA)

Ultraview 1030/1050 Monitors — Service Manual2-19SetupRECORDER NAMES KeyThe RECORDER NAMES key enables you to assign names to each recorder on the sys

Página 64

Ultraview 1030/1050 Monitors — Service Manual2-20PRESELECTED RECORDINGS KeyThe PRESELECTED RECORDINGS key enables you to define which beds and paramet

Página 65 - Maintenance

Ultraview 1030/1050 Monitors — Service Manual2-21SetupTHIS MONITOR includes all the parameters on this monitor.1 Touch the appropriate option.2 Return

Página 66 - Fan Filter

Contents PageiTable of ContentsIntroductionOverview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 67 - Electrical Safety Testing

Ultraview 1030/1050 Monitors — Service Manual2-22FLASHING/STEADY ON KeyThe FLASHING/STEADY ON key selects the mode of operation for the alarm relay (

Página 68 - Definitions

Ultraview 1030/1050 Monitors — Service Manual2-23SetupUSER ACCESS KeyThe USER ACCESS key provides privileged users access to enable certain functions

Página 69 - Chassis Leakage Current Tests

Ultraview 1030/1050 Monitors — Service Manual2-24SERVICE FUNCTIONS MenuThis menu provides the operator with the capability to review system informatio

Página 70 - Functional Checks

Ultraview 1030/1050 Monitors — Service Manual2-25SetupANALOG KeySelecting the ANALOG key displays the battery charge rate menu and its displays. Two d

Página 71

Ultraview 1030/1050 Monitors — Service Manual2-26SITE SURVEY KeyThe SITE SURVEY key is disabled if the radio is not running. This key displays this mo

Página 72 - Capnography

Ultraview 1030/1050 Monitors — Service Manual2-27SetupUpdating Flash PROMNote:Updating Flash PROM is only accomplished by Spacelabs Medical personnel.

Página 73

Ultraview 1030/1050 Monitors — Service Manual2-28Warning:Ensure that the Ethernet wall plate and the shield of the Ethernet connecting cable are bonde

Página 74 - Part Replacement Procedures

Ultraview 1030/1050 Monitors — Service Manual2-29Setup• Verify that the ECG waveform is displayed. 3 Connect a waveform simulator to the pressure inpu

Página 75 - Replacing the Main PCBA

Ultraview 1030/1050 Monitors — Service Manual2-3090310 Rear Panel Ground lug — available for any chassis grounding requirements Antenna connectors

Página 76 - Removing the Chassis Assembly

Ultraview 1030/1050 Monitors — Service Manual2-31SetupConfiguration Options for Wireless Ethernet Option (90310)Figure 2-8: 90310 Wireless Ethernet In

Página 77

Ultraview 1030/1050 Monitors — Service ManualiiAppendix B — Electromagnetic CompatibilityElectromagnetic Emissions. . . . . . . . . . . . . . . . . .

Página 79

Contents3-1TheoryOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1DC Po

Página 80 - Removing the Door

Ultraview 1030/1050 Monitors — Service Manual3-2DC Power SupplyThe power supply is a self-contained device that converts AC power to +18 V for use by

Página 81

Ultraview 1030/1050 Monitors — Service Manual3-3Theory90367/90369 MonitorFigure 3-2 displays the major internal components of the monitor and the basi

Página 82 - Removing the Bezel

Ultraview 1030/1050 Monitors — Service Manual3-4CPU PCBAThe CPU PCBA consists of a CPU subsystem that runs the system software, updates the display, a

Página 83

Ultraview 1030/1050 Monitors — Service Manual3-5TheoryMain CPUThe Ultraview 1030/1050 uses a 25 MHz version of the Motorola MPC860 processor. The foll

Página 84 - Removing the I/O Bay PCBA

Ultraview 1030/1050 Monitors — Service Manual3-6VideoThe Ultraview video is implemented using a Chips and Technologies 65554 VGA controller that resid

Página 85

Ultraview 1030/1050 Monitors — Service Manual3-7TheoryAudioThe Audio subsystem is based on a Crystal CS 4231A Multimedia Codec. This chip is capable o

Página 86 - PCBAs for Option G

Ultraview 1030/1050 Monitors — Service Manual3-8SDLC InterfaceThe SDLC bus is the communications interface to Spacelabs Medical modules, which supply

Página 87

Ultraview 1030/1050 Monitors — Service Manual3-9TheoryThese clocks are used for specific peripherals and are generated at various locations on the PCB

Página 88 - PCBAs for Option H

Contents1-1IntroductionOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 89

Ultraview 1030/1050 Monitors — Service Manual3-10Software ResetTo cause a software reset, the 860 causes a unrecoverable situation by first disabling

Página 90 - System Startup

Ultraview 1030/1050 Monitors — Service Manual3-11TheoryFigure 3-5: Interrupt block diagramwirelesskeyboardaudiomouseISAcontrollersupervisor ICMPC860C1

Página 91 - Boot Menu

Ultraview 1030/1050 Monitors — Service Manual3-12Power Supply SystemFigure 3-6: Power supply block diagramBattery(s) Input/Outputpower supply / batter

Página 92

Ultraview 1030/1050 Monitors — Service Manual3-13TheoryThe power supply has these features: • 7 output DC voltages (+5 V, +3.3 V, +12 V, -12 V, -5 V,

Página 93 - Extended Diagnostics

Ultraview 1030/1050 Monitors — Service Manual3-14Battery ChargersThe battery chargers have the following features:• Front panel LED battery charging/f

Página 94 - Diagnostic Menus

Ultraview 1030/1050 Monitors — Service Manual3-15TheoryNote:The 90367/90369 system contains two identical battery charging systems, one for each batte

Página 95

Ultraview 1030/1050 Monitors — Service Manual3-16EMI ReductionThe CPU PCBA includes several design features intended to reduce EMI effects. The power

Página 96 - MEMORY MENU

Ultraview 1030/1050 Monitors — Service Manual3-17TheoryMAIN Ultraview Care Network SUBSYSTEMS and TASKS:The SDLC subsystem sends packets from the modu

Página 97

Ultraview 1030/1050 Monitors — Service Manual3-18Touchscreen PCBAThe touchscreen PCBA consists of an array of infrared emitters and detectors that sur

Página 98 - Diagnostic LED

Ultraview 1030/1050 Monitors — Service Manual3-19Theory17 gnd Input18 FPSW+ Inputfront panel switch19 FPSW- Output20 gnd Input21 EMITCUR Input Drive v

Página 99 - Error Log

Ultraview 1030/1050 Monitors — Service Manual1-2Figure 1-1: Ultraview 1030/1050 monitors (with Wireless Ethernet Option)Monitor Hardware OptionsThe fo

Página 100

Ultraview 1030/1050 Monitors — Service Manual3-20I/O Bay PCBAGeneralThe basic I/O bay PCBA connects the rear external connectors to the CPU PCBA via t

Página 101 - Troubleshooting

Ultraview 1030/1050 Monitors — Service Manual3-21Theoryremote displayconnectorDB 9-style, 15 pin, female1- ANALOG RED2- ANALOG GREEN3- ANALOG BLUE4- N

Página 102

Ultraview 1030/1050 Monitors — Service Manual3-22Recorder Interface PCBAThe Ultraview 1030/1050 system’s internal recorder is supported by a separate

Página 103 - System Troubleshooting

Ultraview 1030/1050 Monitors — Service Manual3-23TheoryICB (Interconnect PCBA)The interconnect PCBA connects the CPU PCBA to various other parts of th

Página 104 - Touchscreen Troubleshooting

Ultraview 1030/1050 Monitors — Service Manual3-2421 B122 B223 B324 B425 B5 Blue Data MSB26 GND27 ENABSettle the horiz. display position28 VCC +5 V29 V

Página 105 - SDLC Output Troubleshooting

Contents4-1MaintenanceOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Página 106

Ultraview 1030/1050 Monitors — Service Manual4-2• Remove excess liquid from the cloth and squeeze dry.• Wipe exposed surfaces of the equipment and cab

Página 107 - Battery Troubleshooting

Ultraview 1030/1050 Monitors — Service Manual4-3MaintenanceCapnography SensorsThe capnography sensor and cable should be cleaned routinely, especially

Página 108 - Recorder Troubleshooting

Ultraview 1030/1050 Monitors — Service Manual4-4DefinitionsClassification — IEC/EN/UL 60601-1 Safety standard designation for the class of equipment a

Página 109

Ultraview 1030/1050 Monitors — Service Manual4-5MaintenanceWarning:Before starting safety tests, ensure that no patient is connected to the device und

Página 110

Ultraview 1030/1050 Monitors — Service Manual1-3IntroductionExample:A 90369-XFQU04 (where “X” represents the language designator) is an Ultraview 1050

Página 111

Ultraview 1030/1050 Monitors — Service Manual4-6Preventive MaintenanceA qualified Spacelabs Medical repair representative or hospital biomedical techn

Página 112

Ultraview 1030/1050 Monitors — Service Manual4-7MaintenanceVerifying Monitor FunctionsNote:To begin the test, make sure no modules are installed and t

Página 113 - Capnography Troubleshooting

Ultraview 1030/1050 Monitors — Service Manual4-8CapnographyCalibrationRequired Equipment:•O2 sensor (P/N 015-0285-00)*•O2 cable (P/N 175-0877-01 or 17

Página 114 - Does Not Display

Ultraview 1030/1050 Monitors — Service Manual4-9Maintenance6 Attach the O2 sensor to the O2 cable. Verify that the O2 numeric zone appears and display

Página 115 - Check Reads Low

Ultraview 1030/1050 Monitors — Service Manual4-10Part Replacement ProceduresTools/Test Equipment Needed• Anti-static mat with wrist strap• #1 and #2 P

Página 116 - Problem Messages

Ultraview 1030/1050 Monitors — Service Manual4-11MaintenanceReplacing the Main PCBA1 Remove the bezel as previously described.2 With the monitor stand

Página 117

Ultraview 1030/1050 Monitors — Service Manual4-12Removing the Chassis AssemblyThe chassis assembly includes the sheet metal Interconnect PCBA, and the

Página 118

Ultraview 1030/1050 Monitors — Service Manual4-13MaintenanceFigure 4-5: Chassis out of the case7 To lift the chassis out of the case, place your left

Página 119 - Field Replaceable Parts List

Ultraview 1030/1050 Monitors — Service Manual4-14Replacing the Fan/Battery Contact Assembly1 Remove the chassis assembly from the housing as previousl

Página 120

Ultraview 1030/1050 Monitors — Service Manual4-15MaintenanceReplacing the Interconnect PCBA and Frame1 Remove the chassis assembly from the housing as

Página 122

Ultraview 1030/1050 Monitors — Service Manual4-16Replacing/Aligning the Module DoorRemoving the Door1 Remove the two nuts fastening the hinge to the w

Página 123 - Drawings

Ultraview 1030/1050 Monitors — Service Manual4-17MaintenanceReplacing the I/O PCBA — Option FRemoving the Option Panel1 Remove the five 6-32 × 1/2 inc

Página 124

Ultraview 1030/1050 Monitors — Service Manual4-18Display or Touchscreen Assembly ReplacementRemoving the BezelFigure 4-10: Bezel removalNote:The handl

Página 125 - Appendix A — Glossary

Ultraview 1030/1050 Monitors — Service Manual4-19MaintenanceCaution:Never pull on the flex cables. This may damage the cable end or the connector.1 Di

Página 126

Ultraview 1030/1050 Monitors — Service Manual4-20Capnograph with Ethernet Removal/Disassembly — Option G or HRemoving the I/O Bay PCBAFigure 4-12: Eth

Página 127

Ultraview 1030/1050 Monitors — Service Manual4-21Maintenance3 Remove the six jackscrews and the knurled nuts (two each) not shown. The knurled nuts ar

Página 128

Ultraview 1030/1050 Monitors — Service Manual4-223 Remove the four flathead screws under the label.4 Unplug the cable between J2 and J4 taking appropr

Página 129

Ultraview 1030/1050 Monitors — Service Manual4-23MaintenanceRemoving the EMI Interface PCBA for Option HFigure 4-17: PCBA support plate and EMI interf

Página 130

Ultraview 1030/1050 Monitors — Service Manual4-24Removing the CO2 PCBAs for Option HFigure 4-18: CO2/SDLC PCBA and sensor interface PCBA for option H1

Página 131 - Compatibility

Contents5-1TroubleshootingOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 132 - Electromagnetic Immunity

Contents2-1SetupUnpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Assembl

Página 133

Ultraview 1030/1050 Monitors — Service Manual5-2System StartupWhen the unit is first powered ON, the following three window system startup screen is d

Página 134 - Mitigation

Ultraview 1030/1050 Monitors — Service Manual5-3TroubleshootingBoot MenuThe Boot menu allows access to several basic configuration menus and functions

Página 135 - Appendix C — Symbols

Ultraview 1030/1050 Monitors — Service Manual5-4After initiating the Boot menu, the following menu appears.Figure 5-3: Boot menuThese keys are present

Página 136

Ultraview 1030/1050 Monitors — Service Manual5-5TroubleshootingPower ON DiagnosticsPower-ON diagnostic tests verify system hardware integrity during e

Página 137 - 12,200 m

Ultraview 1030/1050 Monitors — Service Manual5-6For detailed information on the extended diagnostics and how to run them, refer to Diagnostic Menus on

Página 138

Ultraview 1030/1050 Monitors — Service Manual5-7TroubleshootingTouch the i in the Diagnostic main menu to present the Individual Diagnostic menu:Figur

Página 139

Ultraview 1030/1050 Monitors — Service Manual5-8Touch the m key in the Individual Diagnostic menu to present the Memory menu:Figure 5-6: Memory menuTh

Página 140 - ETL Laboratory Approved

Ultraview 1030/1050 Monitors — Service Manual5-9TroubleshootingTouch the v key in the Individual Diagnostic menu to present the Video menu:Figure 5-7:

Página 141

Ultraview 1030/1050 Monitors — Service Manual5-10Touch the t key in the Individual Diagnostic menu to present the Touchscreen menu:Figure 5-8: Touchsc

Página 142

Ultraview 1030/1050 Monitors — Service Manual5-11TroubleshootingError LogFor troubleshooting intermittent problems, the monitor maintains a text error